A N A L Y Z E R 90
Clinical Chemistry Photometer



Reliable and robust semi-automated clinical chemistry 10 mm Flow Cell photometer. Flowcell with Peltier controlled temperature. Standard configuration with 7 optical filters, wave length 340-400/700 mm. Mono and bichromatic reading systems. Its operating software runs and stores up to 200 tests. Reading system: End-Point, Fixed-Time, and Kinetic; data processing with different calculation methods.
TECHNICAL FEATURES
Accuracy ± 1% from 0.000 to 1.500 OD
± 2% from 1.500 to 3.000 OD
capability range 0.000 - 4.000 OD
Stability 0.01 OD after 8 hours
Light source halogen lamp
Optical filters 340, 405, 492, 510, 546, 578, 630
Detector silicon photodiode
Reading path 10 mm
Flow-cell 80 µl
Flow-cell temp room temperature 25 – 30 – 37°C ± 0.1
Aspiration automated sample aspiration through a precision peristaltic pump
Aspirating volume minimum 350 µl
Memory capacity 200 tests and curves; the last 300 readings to be printed and sent to the serial port
Reading system mono and bi-chromatic
Fast reading for kinetic and fixed time
Calculation mode optical density, concentration, endpoint against standard of factor, fixtime against standard of factor, kinetic, non linear multi-standard curve up to 8 points, blank sample, blank reagent
Calibration from 2 to 8 points curve
Calendar timer associates time with the results; timer function
Keypad 17 buttons
Display 2 lines with 16 columns
Printer 40 columns with thermal paper
Interface RS232 serial port
Warm-up time 15 minutes
Power 230/115 VAC, 50-60 Hz, 110 W
Measurements cm 42 x 34 x 15 h
Weight 9 Kg