TECHNICAL FEATURES
  
Flow-rates range0.25 - 25 l/min
Precision1% from displayed value
Accuracy2% from displayed value
Flow-rate constancy0.5% from displayed value
Self-testsleak test; pneumatic calibration test
Alarmsauto-alarms preventing malfunctioning of the sampler. Additional alarms can be set by the user in order to show unwanted operative conditions
Data available during the samplingInlet flow-rate
Flow-rate at standard conditions
External temperature
On-line pressure drop
Data available after the samplingStandard sampled volume
Initial and final pressure drop
Atmospheric pressure
Sampling status
Internal memory four operative programs which can be modified by the user; the data of the last 32 samplings are stored in memory
InterfacesBack-lighted liquid cristals display
Sensitive membrane keypad
20 columns thermal printer
RS 232 which allows to run the system from a remote computer
RS 485 which links the sampler to other units
Connection plug to external sensors and for driving external lines
PeripheralsADS7 runs denuders and filter-packs systems
PMS runs membrane filter systems
VOCS samples volatile organic compounds on adsorbing traps
M16 connects up to 16 pneumatics lines to Sequair 92/A
METEO associates each sampling line to a specified meteorological condition
STACK runs automatic sampling in isokinetic conditions
Supply220Vac 50/60Hz 500W
Dimensionscm 43 x 60 x 26
WeightKg 37